Performance Characteristics
- Entire test performed in-house
As of March 2016, 6.36 Mb of sequence (83 genes, 1557 exons) generated in our lab was compared between Sanger and NextGen methodologies. We detected no differences between the two methods. The comparison involved 6400 total sequence variants (differences from the reference sequences). Of these, 6144 were nucleotide substitutions and 256 were insertions or deletions. About 65% of the variants were heterozygous and 35% homozygous. The insertions and deletions ranged in length from 1 to over 100 nucleotides. In silico validation of insertions and deletions in 20 replicates of 5 genes was also performed. The validation included insertions and deletions of lengths between 1 and 100 nucleotides. Insertions tested in silico: 2200 between 1 and 5 nucleotides, 625 between 6 and 10 nucleotides, 29 between 11 and 20 nucleotides, 25 between 21 and 49 nucleotides, and 23 at or greater than 50 nucleotides, with the largest at 98 nucleotides. All insertions were detected. Deletions tested in silico: 1813 between 1 and 5 nucleotides, 97 between 6 and 10 nucleotides, 32 between 11 and 20 nucleotides, 20 between 21 and 49 nucleotides, and 39 at or greater than 50 nucleotides, with the largest at 96 nucleotides. All deletions less than 50 nucleotides in length were detected, 13 greater than 50 nucleotides in length were missed. Our standard NextGen sequence variant calling algorithms are generally not capable of detecting insertions (duplications) or heterozygous deletions greater than 100 nucleotides. Large homozygous deletions appear to be detectable. The PGxome test detects most larger deletions and duplications including intragenic CNVs and large cytogenetic events; however aberrations in a small percentage of regions may not be accurately detected due to sequence paralogy (e.g., pseudogenes, segmental duplications), sequence properties, deletion/duplication size (e.g., 1-3 exons vs. 4 or more exons), and inadequate coverage. In general, sensitivity for single, double, or triple exon CNVs is ~70% and for CNVs of four exon size or larger is >95%, but may vary from gene-to-gene based on exon size, depth of coverage, and characteristics of the region.
Proficiency Testing (PT)
- Is proficiency testing performed for this test?
HelpWhether the test undergoes periodic internal or external evaluation of the accuracy of test results. Lab-provided.
- Yes
- Method used for proficiency testingHelpMethod used for external or internal evaluation of the accuracy of test results. Lab-provided.
- Intra-Laboratory
- FDA Category Designation
- FDA exercises enforcement discretion