Approaches for Successful Outcomes with Continuous Glucose Monitoring

, MD, FACE, FACP.

Author Information and Affiliations

Published: August 2018.

Successful management of diabetes requires regular monitoring of glucose levels for all patients (1), with greater frequency recommended for those with type 1 or insulintreated type 2 diabetes (2). However, daily self-monitoring of blood glucose (SMBG) may be painful, inconvenient, costly, and difficult to maintain. In 2017, a study using Cloud-based analysis software revealed that rates of glucose monitoring in Europe and North America ranged between 2.7 and 4.4 times/day in people with any type of diabetes (3). Individuals who perform SMBG typically focus on pre-meal or bedtime glucose levels, obtaining only a static snapshot of points in time. In the past two decades, continuous glucose monitoring (CGM), using subcutaneous sensors to measure interstitial glucose levels, has emerged to provide a better understanding of glucose trends and patterns.

Real-time CGM devices, alone or integrated into insulin pump systems, display data continuously and provide alerts and alarms for current and impending hypoglycemia and hyperglycemia (4,5). In 2017, a novel factory-calibrated, sensor-based system for daily use by people with diabetes, the FreeStyle Libre (Abbott, Alameda, CA), became available in the United States (6). The Libre’s technology has been alternately referred to as flash CGM (FCGM) and intermittently scanned CGM because continuous data are viewable to the user only when a dedicated reader is scanned (or “flashed”) over the sensor (7). The FreeStyle Libre has no alarms, but a distinct audible tone is provided when alerts to perform fingerstick testing are displayed. This may occur during scanning when glucose is <70 or >240 mg/dL, projected to be <70 or >240 mg/dL, “hi” or “lo,” projected to be “hi” or “lo,” or when glucose is rapidly changing or no trend arrow displays. A built-in glucose meter and individually foil-packed glucose strips facilitate measurement of glucose levels in these situations (8). The FreeStyle Libre does not require fingerstick testing for calibration.

The accuracy of CGM systems has improved over time, and presently, several available systems are approved as tools for making treatment decisions. All have similar accuracy at glucose levels >80–200 mg/dL. However, in the hypoglycemic range, the FreeStyle Libre system is not as accurate as in the euglycemic range (9), and sensor readings should be confirmed with blood glucose measurements. All personal CGM systems provide current glucose trend arrows. Because the meaning of the arrows is systemspecific (see the article on p. 8 of this compendium), health care providers must learn the differences among the devices and guide patients based on each one’s ecosystem for therapy adjustment. In the United States, the Dexcom G5 Mobile and FreeStyle Libre systems are approved by Medicare for beneficiaries with diabetes who use intensive insulin therapy (three or more injections per day), perform fingerstick glucose testing four times per day, and require frequent adjustment in therapy (1013).

Outcomes of CGM in people with type 1 and insulintreated type 2 diabetes are reviewed in the article on p. 3 of this compendium (1418). Results in more heterogeneous groups with type 2 diabetes show variable effectiveness and acceptability (19). With high adherence to CGM, increased physical activity, reduced calorie intake, and decreased body weight were observed. This is consistent with findings that increased frequency of SMBG and CGM correlate with positive outcomes in type 1 diabetes (1418,20). Benefits of FCGM have also been demonstrated in the IMPACT and REPLACE studies for type 1 and type 2 diabetes, respectively, with overall time spent in hypoglycemia reduced by 38% (type 1 diabetes) and 43% (type 2 diabetes) (21,22). Previously, improved glycemic control for up to 1 year was observed in patients not on prandial insulin using intermittent real-time CGM (23).

For some people, fatigue from alerts and alarms may thwart improved glucose outcomes with real-time CGM (24,25). FCGM, which has no alarms and sounds a distinct audible tone during scanning when alerts to perform fingerstick testing are displayed, offers a viable alternative; in studies of FCGM to date, patient satisfaction and adherence have been high (21,22).

Frequency of Looking at Receiver/Reader Data

It is difficult to quantify how often users check real-time CGM data during the day and night, either actively before insulin dosing or passively when alerted to hypoglycemia or hyperglycemia; however, a reasonable estimate is at least 4–12 times/day, including before meals, at bedtime, for physical activity, and in response to alerts and alarms. Quantification of FCGM is easier, as the number of scans per day is provided on the reader and available when uploading data to the LibreView software. In two recent studies using FCGM, the average number of scans per day was reported to be 15 for type 1 diabetes and 8 for type 2 diabetes patients (21,22). A recent analysis of FCGM in >50,000 users worldwide provided additional insight into real-world experience. In this report, the number of scans per day positively correlated with glycemic outcomes, with less time spent in hypoglycemia and hyperglycemia and more time spent in range with increasing number of scans per day. The number of scans per day ranged from 4.4 (every 5.4 hours) to 48 (every 30 minutes) (26). Health care providers should address the ideal frequency of scanning on an individual basis and modify recommendations based on each patient’s treatment regimen and needs.

Patient Selection

Careful patient selection is important when recommending CGM therapy. The article on p. 11 of this compendium provides more details about appropriate candidates. Successful outcomes will depend in large part on a person’s trust in the system, willingness to calibrate the system per product specifications, the number of times the person scans or looks at the system, and the type of intervention plan set up with the health care provider. Guidelines for CGM patient selection have been developed by professional societies and other expert forums (2729).

Regardless of baseline A1C or the degree of glucose variability at CGM initiation, users should be willing to check or scan their device on a near-daily basis to realize the greatest benefit (1418,21,22). Users should also understand the concept of interstitial fluid versus capillary blood glucose measurements and calibration procedures for systems that require calibration. Of note, for real-time CGM systems, setting alerts and alarms with realistic expectations is essential to avoid alarm fatigue (24). Establishing a plan for sick-day or illness management with CGM is greatly encouraged. Dexcom CGM users should also consider taking advantage of the Share feature, which allows a “follower” (person chosen by the user, such as a parent, family member, or friend) to receive CGM information on a smartphone. Such data, including alerts and alarms, would allow the recipient to potentially assist the user if necessary, such as in the event of hypoglycemia. Educating family members or other caretakers about the CGM system and, for older adults, ensuring that they can see or hear the alerts and alarms, are also fundamental to successful outcomes.

Patient and Provider Education

The importance of patient and provider education cannot be overemphasized. People with diabetes should receive training on the meaning of the messages displayed on their system reader. Additionally, setting procedures for alerts and alarms, individualizing trend arrow-based treatment decisions, and reinforcing the dangers of insulin stacking (administering insulin while the previous dose is still active) are crucial to promoting adherence to CGM and improving glycemic outcomes. Although training videos are provided by manufacturers, they should not be seen as a substitute for in-depth patient education, especially when initiating CGM. Without appropriate training, CGM users may not be able to take full advantage of the information provided. Additionally, because most health care providers lack training in the interpretation of CGM data, including retrospective analysis during office visits, the availability of educational resources for health care providers is essential to achieving positive outcomes.

Hypoglycemia

Hypoglycemia is a serious concern for people with diabetes and the major limiting factor in achieving glycemic targets with intensive management. Hypoglycemia risks and frequency are well established in people with type 1 diabetes (30,31), and even self-reported severe hypoglycemia is associated with a 3.4-fold increased risk of death (32). However, hypoglycemia frequency may be underestimated in people with type 2 diabetes; whether on insulin or other antihyperglycemic regimens, these individuals can also have hypoglycemia unawareness (33). CGM technology has been demonstrated to reduce hypoglycemia frequency and hypoglycemia unawareness in people with diabetes (1418,21,22,3437).

When recommending CGM to patients, the presence of hypoglycemia unawareness should be evaluated and discussed to aid in selecting the most appropriate system.

Interpretation of Data

Interpretation of data for users and providers still lacks a cohesive approach. When reviewing data, time spent in the various ranges (<54, <70, 70–180, >180, and >250 mg/ dL) as well as coefficient of variation should be addressed. This can be accomplished with Ambulatory Glucose Profile reports available in various software platforms (Dexcom Clarity, Glooko-Diasend, Tidepool, Medtronic CareLink, and LibreView) (3842). Nocturnal hypoglycemia should be addressed first, with intervention to reduce its severity and frequency. Subsequently, fasting glucose levels should be evaluated, and modifications to basal insulin doses or insulin-pump basal rate settings should be implemented. CGM systems offer great advantage in identifying prandial glycemic excursions; in such cases, mealtime insulin doses and the timing of the boluses should be addressed by instructing patients to monitor glucose before and 2–4 hours after meals to better understand glucose fluctuations and make appropriate regimen adjustments. Recently, use of FCGM was associated with a significant increase in delivering bolus insulin 15–20 minutes in advance of meals (compared to immediately before or after meals) (36). People using insulin should be cautioned against making frequent dose changes in response to above-target post-meal CGM readings on their display tools because insulin stacking is a well-known and avoidable cause of hypoglycemia.

Trend Arrows

As mentioned previously, CGM systems feature trend arrows that provide information on the predicted change of glucose levels over a specific time period. It is important to be aware that the arrows correspond to different rates of glucose change depending on the brand of CGM system. Training users on the meaning of arrows displayed on their particular device will ensure that they take appropriate actions guided by the specific system they use. (See the table on p. 9 of this compendium for details about each available system.) Although approaches to adjusting insulin doses on the basis of trend arrows are frequently discussed and constitute an important aspect of therapeutic CGM, proposed methods have yet to be validated in randomized controlled trials (4347). The most recently published method is specific to the Dexcom G5 Mobile (47). With this in mind, it is imperative that health care providers take an individualized approach when applying trend arrows to treatment decisions (Table 1).

TABLE 1

TABLE 1

Published Trend Arrow Methods for Insulin Dose Adjustment

In view of the aforementioned caveats, the published recommendations made with respect to the Dexcom G5 arrows cannot be extrapolated to the FreeStyle Libre system. However, using a correction factor for trend arrow–guided insulin dose adjustments in combination with mealtime dose calculations based on an insulin-to-carbohydrate ratio offers an advantage over other published methods in that it allows for personalized dose calculations based on insulin sensitivity. A proposed working algorithm for the FreeStyle Libre incorporating this concept is shown in Table 2, and sample dose adjustments are provided in Box 1. The rationale for this non-validated method is to be more aggressive when the reader displays one up-trending arrow because the predicted glucose change over time with the FreeStyle Libre in reality could be much higher than 2 mg/ dL per minute. Moreover, readings in the hypoglycemic range must be confirmed with fingerstick glucose testing because of potential sensor inaccuracy; when glucose is rapidly falling and sensor glucose levels are <100 mg/dL, additional rapid-acting carbohydrate intake (15–30 g) should be considered in the pre-meal period. In the postprandial period, insulin dosing based on trend arrows should be calculated 4 hours after an insulin dose, although this recommendation may change with the recent introduction of insulin analogs with a faster onset of action (48).

TABLE 2

TABLE 2

Suggested Insulin Dose-Adjustment Algorithm for Free Style Libre Trend Arrows

Box Icon

BOX 1

Examples of Trend Arrow–Guided Insulin Dose Adjustments with FreeStyle Libre

Before using trend arrows for insulin dose adjustments, patients should first become familiar with their CGM systems. Because users will work within their own CGM ecosystem, education should be targeted toward patients’ specific device. Although there are no head-to-head trials, all available CGM systems have decreased accuracy in the hypoglycemic range (9,49,50), but in the United States, only the FreeStyle Libre label requires confirmation with fingerstick testing in this range. Thus, individualized guidance may be needed on using trend arrows for insulin dose adjustments when glucose levels are <70 mg/dL. This is especially true in older adults with diabetes, who are at highest risk for severe hypoglycemia and hypoglycemia unawareness (51). For certain high-risk geriatric patients, use of trend arrows for insulin dose changes with the FreeStyle Libre should be assessed on a case-by-case basis, and it may be prudent to confirm glucose levels via fingerstick glucose testing when adjusting insulin per instructions from the reader display; in the setting of rapid glucose changes toward hypoglycemia, fingerstick glucose checks should be strongly recommended for safety, especially when the “check blood glucose” symbol appears on the reader display. Of note, in a small cohort of nursing home residents with type 2 diabetes, FCGM overestimated hypoglycemia, with 51.4% of the interstitial glucose readings <70 mg/dL being falsely low compared to capillary blood glucose levels (52). More studies of FCGM in geriatric populations are needed.

Non-Insulin-Using Patients

Formal studies of CGM in non-insulin-using patients on antihyperglycemic regimens are scant. For such people, the main goal of CGM should be to achieve target fasting glucose levels and decrease postprandial glycemic excursions with appropriate regimen adjustments, meal quality modifications, and lifestyle interventions (2,53). For people with type 2 diabetes, especially those on sodium– glucose cotransporter 2 inhibitors or glucagon-like peptide 1 receptor agonists, CGM therapy could have significant educational and therapeutic benefits, with the additional advantage of obtaining data relatively painlessly. Developing a plan to review CGM trends in the postprandial period, whether by setting up alerts or by instructing users to scan their FCGM at specific times after meals, can convey the dynamics of post-meal glucose fluctuations to patients and guide providers in personalizing diabetes regimen adjustments based on data accumulated between visits. More studies are needed to address the potential value of FCGM in this clinical setting.

Ideally, all systems should display glucose data on a mobile app so that users do not have to carry multiple tools; this strategy would potentially contribute to acceptance and increase uptake of CGM use in non-insulin-using people with diabetes.

Best Practice for Exercise

Standards of care recommend that most adults with type 1 or type 2 diabetes engage in daily physical activity, allowing no more than two consecutive days without activity (2,53).

However, blood glucose responses to physical activity, especially in individuals with type 1 diabetes, can be highly variable depending on many factors, including the type and timing of activity, previous food ingestion, and level of insulin on board. Adjustments in medication doses and carbohydrate intake are often required to maintain adequate glucose levels during and after physical activity (54,55).

CGM can be of benefit by providing users with glucose trend data at any time, thereby decreasing fear of exercise induced hypoglycemia. This is especially relevant when sensor glucose levels are trending toward hypoglycemia. However, very few studies have addressed this issue, and the accuracy of CGM has not been fully validated with different types of exercise. For example, intermittent high intensity interval exercise is associated with metabolic changes (e.g., changes in pH, microcirculation, and oxygen tension) that may potentially interfere with CGM accuracy. In a small study using the Dexcom G4 Platinum system, the accuracy was comparable during continuous moderate and intermittent high-intensity exercise during a cycling session (56). Similarly, the accuracy of real-time CGM (Medtronic Guardian REAL-Time) measured at various prescribed workloads was acceptable for all types of exercises with the exception of continuous high-intensity exercise, where lower accuracy was detected (57). A recent survey of 502 adults from the T1D Exchange’s online patient community, 276 of whom were using CGM either alone or in combination with CSII, showed that, although most respondents adjusted carbohydrate intake and insulin doses around exercise, the majority still reported experiencing hypoglycemia after exercise and having significant difficulties with blood glucose control around exercise (58).

Using CGM trend arrows adds another level of complexity. Although trend arrows are helpful in determining the direction of glucose during exercise and guiding users in the consumption of additional carbohydrate to prevent or reduce hypoglycemia, using them to adjust insulin doses during and after exercise is more challenging. As previously suggested (47), users should be conservative when adjusting their insulin doses before exercise and should refrain from increasing their insulin dose in the presence of up-trending arrows during the active period. In the postexercise period, as previously recommended by Riddell et al. (55), close glucose monitoring is essential. In particular, attention should be paid to the direction of trend arrows in the immediate or even late-post-exercise period to correct impending hypoglycemia with rapid-acting carbohydrate intake, if indicated. Trend arrows may also signal the need for additional carbohydrate or an insulin dose reduction at bedtime to avoid nocturnal hypoglycemia, especially when exercise takes place in the late afternoon or evening (55). It is hoped that additional studies of newer CGM systems and their accuracy in response to various exercise protocols will better define best practice for the use of CGM trend arrows during physical activity.

Dualities of Interest

I.B.H. has served as a consultant to Abbott Diabetes Care, Adocia, Bigfoot, and Roche. His institution has received research grant support from Medtronic.

T.B. has served on advisory boards of Bayer Health Care, Boehringer Ingelheim, DreaMed Diabetes, Eli Lilly, Medtronic, Novo Nordisk, and Sanofi. His institution has received research grant support, with receipt of travel and accommodation expenses in some cases, from Abbott Diabetes Care, Diamyd, GluSense, Medtronic, Novo Nordisk, Sandoz, and Sanofi. He has received honoraria for participating on the speakers bureaus of Bayer Health Care, Eli Lilly, Medtronic, Novo Nordisk, Roche, and Sanofi. He owns stock in DreaMed Diabetes.

A.L.P. has served on advisory boards for Abbott Diabetes Care, Becton Dickinson, Bigfoot, Boehringer Ingelheim, Eli Lilly, Lexicon, Livongo, Medscape, Merck, Novo Nordisk, OptumHealth, Sanofi, and Science 37. She has received research grant support from Dexcom and Mannkind. She participates on a speakers bureau for Novo Nordisk.J.J.C. participates in speakers bureaus for Janssen, Merck, Novo Nordisk, and Sanofi.

G.A. has served as a consultant and on a steering committee for Dexcom and on an advisory board for Novo Nordisk, and her institution has received research grant support from AstraZeneca and Novo Nordisk.

R.M.B.’s institution has received payment for his services as a research investigator, consultant, or advisory board member for Abbott Diabetes Care, Becton Dickinson, Boehringer Ingelheim, Bristol-Myers Squibb/AstraZeneca, Dexcom, Eli Lilly, Hygieia, Johnson & Johnson, Medtronic, Merck, Novo Nordisk, Roche, Sanofi, and Takeda. R.M.B. has inherited Merck stock, volunteers for the American Diabetes Association and JDRF, and receives funding from the National Institutes of Health for diabetes technology research.

Acknowledgments

Writing support services for this compendium were provided by Carol Verderese of The Diabetes Education Group in Lakeville, CT. Editorial and project management services were provided by Debbie Kendall of Kendall Editorial in Richmond, VA.

Author Contributions

All authors researched and wrote their respective section(s). Lead author I.B.H. reviewed all content and is the guarantor of this work.

REFERENCES

1.
Schnell O, Alawi H, Battelino T, et al. Self-monitoring of blood glucose in type 2 diabetes: recent studies. J Diabetes Sci Technol 2013;7:478–488 [PMC free article: PMC3737650] [PubMed: 23567007]
2.
American Diabetes Association. Glycemic targets. Sec. 6 in Standards of Medical Care in Diabetes—2018. Diabetes Care 2018;41(Suppl. 1):S55–S64 [PubMed: 29222377]
3.
Glooko. Annual diabetes report 2017. Available from www​.glooko.com/resource​/2017-annual-diabetes-report. Accessed 3 March 2018
4.
Medtronic. Outsmart your diabetes: New! The Guardian Connect CGM system. Available from www.medtronicdiabetes.com/products/ guardian-connect-continuous-glucose-monitoring-system. Accessed 27 April 2018
5.
Dexcom. What is (CGM) continuous glucose monitoring? Available from www​.dexcom.com/continuous-glucose-monitoring. Accessed 27 April 2018
6.
Abbott. Discover the FreeStyle Libre system. Available from www​.freestylelibre.us​/about-system-overview. Accessed 27 April 2018
7.
Petrie JR, Peters AL, Bergenstal RM, Holl RW, Fleming GA, Heinemann L. Improving the clinical value and utility of CGM systems: issues and recommendations: a joint statement of the European Association for the Study of Diabetes and the American Diabetes Association Diabetes Technology Working Group. Diabetes Care 201740:1614–1621 [PubMed: 29070577]
8.
Abbott. Accuracy meets affordability: the FreeStyle Precision Neo blood glucose meter and test strips are the solution to your need for an affordable system that delivers results with proven accuracy. Available from www​.myfreestyle.com/neo. Accessed 27 April 2018
9.
Fokkert MJ, van Dijk PR, Edens MA, et al. Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care 2017;5:e000320 [PMC free article: PMC5316912] [PubMed: 28243449]
10.
Centers for Medicare & Medicaid Services. CMS rulings: ruling no. [CMS-1682-R]. Available from www​.cms.gov/Regulations-andGuidance​/Guidance​/Rulings/Downloads/CMS1682R.pdf. Accessed 8 April 2018
11.
Noridian Healthcare Solutions. Coding and coverage: therapeutic continuous glucose monitors (CGM). Available from med. noridianmedicare.com/web/jddme/policies/dmd-articles/coding-andcoverage-therapeutic-continuous-glucose-monitors. Accessed 8 April 2018
12.
Dexcom. Dexcom G5 CGM system is now covered under Medicare. Available from www​.dexcom.com/medicare-coverage. Accessed 8 April 2018
13.
Tucker ME. Medicare to cover FreeStyle Libre glucose monitoring system. Available from www​.medscape.com/viewarticle/890879. Accessed 8 April 2018
14.
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008;359:1464–1476 [PubMed: 18779236]
15.
Bergenstal RM, Tamborlane WV, Ahmann A, et al; STAR 3 Study Group. Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med 2010;363:311–320 [PubMed: 20587585]
16.
Beck RW, Riddlesworth T, Ruedy K, et al; DIAMOND Study Group. Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA 2017;317:371–378 [PubMed: 28118453]
17.
Lind M, Polonsky W, Hirsch IB, et al. Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections: the GOLD randomized clinical trial. JAMA. 2017;317:379–387 [PubMed: 28118454]
18.
Beck RW, Riddlesworth TD, Ruedy K, et al; DIAMOND Study Group. Continuous glucose monitoring versus usual care in patients with type 2 diabetes receiving multiple daily insulin injections: a randomized trial. Ann Intern Med 2017;167:365–374 [PubMed: 28828487]
19.
Taylor PJ, Thompson CH, Brinkworth GD. Effectiveness and acceptability of continuous glucose monitoring for type 2 diabetes management: a narrative review. J Diabetes Investig 2018. Epub ahead of print (DOI: 10.1111/jdi.12807) [PMC free article: PMC6031515] [PubMed: 29380542]
20.
Miller KM, Beck RW, Bergenstal RM, et al; T1D Exchange Clinic Network. Evidence of a strong association between frequency of self-monitoring of blood glucose and hemoglobin A1c levels in T1D Exchange clinic registry participants. Diabetes Care 2013;36:2009–2014 [PMC free article: PMC3687326] [PubMed: 23378621]
21.
Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet 2016;388:2254–2263 [PubMed: 27634581]
22.
Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Flash glucose-sensing technology as a replacement for blood glucose monitoring for the management of insulin-treated type 2 diabetes: a multicenter, open-label randomized controlled trial. Diabetes Ther 2017;8:55–73 [PMC free article: PMC5306122] [PubMed: 28000140]
23.
Vigersky RA, Fonda SJ, Chellappa M, Walker MS, Ehrhardt NM. Short- and long-term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care 2012;35:32–38 [PMC free article: PMC3241321] [PubMed: 22100963]
24.
Shivers JP, Mackowiak L, Anhalt H, Zisser H. ‘‘Turn it off!’’: diabetes device alarm fatigue considerations for the present and the future. J Diabetes Sci Technol 2013;3:789–794 [PMC free article: PMC3869147] [PubMed: 23759412]
25.
Wong JC, Foster NV, Maahs DM, et al; T1D Exchange Clinic Network. Real-time continuous glucose monitoring among participants in the T1D Exchange clinic registry. Diabetes Care 2014;37:2702–2709 [PMC free article: PMC4392936] [PubMed: 25011947]
26.
Dunn TC, Xu Y, Hayter G, Ajjan RA. Real-world flash glucose monitoring patterns and associations between self-monitoring frequency and glycaemic measures: a European analysis of over 60 million glucose tests. Diabetes Res Clin Pract 2018;137:37–46 [PubMed: 29278709]
27.
Peters AL, Ahmann AJ, Battelino T, et al. Diabetes technology—continuous subcutaneous insulin infusion therapy and continuous glucose monitoring in adults: an Endocrine Society clinical practice guideline. J Clin Endocrinol Metab 2016;101:3922–3937 [PubMed: 27588440]
28.
Fonseca VA, Grunberger G, Anhalt H, et al.; Consensus Conference Writing Committee. Continuous glucose monitoring: a consensus conference of the American Association of Clinical Endocrinologists and American College of Endocrinology. Endocr Pract. 2016;22:1008–1021 [PubMed: 27214060]
29.
Adolfsson P, Parkin CG, Thomas A, Krinelke LG. Selecting the appropriate continuous glucose monitoring system: a practical approach. Eur Endocrinol 2018;14:24–29 [PMC free article: PMC5954591] [PubMed: 29922348]
30.
Frier BM. Hypoglycaemia in diabetes mellitus: epidemiology and clinical implications. Nat Rev Endocrinol 2014;10:711–722 [PubMed: 25287289]
31.
Geddes J, Schopman JE, Zammitt NN, Frier BM. Prevalence of impaired awareness of hypoglycaemia in adults with type 1 diabetes. Diabet Med 2008;25:501–504 [PubMed: 18387080]
32.
McCoy RG, Van Houten HK, Ziegenfuss JY, Shah ND, Wermers RA, Smith SA. Increased mortality of patients with diabetes reporting severe hypoglycemia. Diabetes Care 2012;35:1897–1901 [PMC free article: PMC3425008] [PubMed: 22699297]
33.
Gehlaut RR, Dogbey GY, Schwartz FL, Marling CR, Shubrook JH. Hypoglycemia in type 2 diabetes: more common than you think: a continuous glucose monitoring study. J Diabetes Sci Technol 2015;9:999– 1005 [PMC free article: PMC4667336] [PubMed: 25917335]
34.
Reddy M, Jugnee N, El Laboudi A, Spanudakis E, Anantharaja S, Oliver N. A randomized controlled pilot study of continuous glucose monitoring and flash glucose monitoring in people with type 1 diabetes and impaired awareness of hypoglycaemia. Diabet Med 2018;35:483–490 [PMC free article: PMC5888121] [PubMed: 29230878]
35.
Heinemann L, Freckmann G, Ehrmann D, et al. Real-time continuous glucose monitoring in adults with type 1 diabetes and impaired hypoglycaemia awareness or severe hypoglycaemia treated with multiple daily insulin injections (HypoDE): a multicentre, randomised controlled trial. Lancet 2018;391:1367–1377 [PubMed: 29459019]
36.
Dover AR, Stimson RH, Zammitt NN, Gibb FW. Flash glucose monitoring improves outcomes in a type 1 diabetes clinic. J Diabetes Sci Technol 2017;11:442–443 [PMC free article: PMC5478014] [PubMed: 27464754]
37.
Choudhary P, Ramasamy S, Green L, et al. Real-time continuous glucose monitoring significantly reduces severe hypoglycemia in hypoglycemia-unaware patients with type 1 diabetes. Diabetes Care 2013;36:4160–4162 [PMC free article: PMC3836107] [PubMed: 24103902]
38.
Dexcom. Dexcom Clarity diabetes management software. Available from www​.dexcom.com/clarity. Accessed 8 April 2018
39.
Glooko. Diasend is office data uploads made easy. Available from www​.glooko.com/diasend. Accessed 8 April 2018
40.
Tidepool Project. Tidepool: where you can see all of your data in one place. Available from www​.tidepool.org. Accessed 8 April 2018
41.
Medtronic. Welcome to the CareLink personal therapy management software for diabetes. Available from carelink.minimed.com/patient/entry. jsp?bhcp=1. Accessed 8 April 2018
42.
Abbott. Clear, intuitive reports for you and your doctor. Available from pat​.libreview.com. Accessed 8 April 2018
43.
Diabetes Research in Children Network (DirecNet) Study Group. Use of the DirecNet Applied Treatment Algorithm (DATA) for diabetes management with a real-time continuous glucose monitor (the FreeStyle Navigator). Pediatr Diabetes 2008;9:142–147 [PMC free article: PMC2390770] [PubMed: 18221427]
44.
Scheiner G. Practical CGM: Improving Patient Outcomes Through Continuous Glucose Monitoring. 4th ed. Alexandria, VA, American Diabetes Association, 2015
45.
Pettus J, Edelman SV. Recommendations for using real-time continuous glucose monitoring (rtCGM) data for insulin adjustments in type 1 diabetes. J Diabetes Sci Technol 2017;11:138–147 [PMC free article: PMC5375074] [PubMed: 27530720]
46.
Klonoff DC, Kerr D. A simplified approach using rate of change arrows to adjust insulin with real-time continuous glucose monitoring. J Diabetes Sci Technol 2017;11:1063–1069 [PMC free article: PMC5951054] [PubMed: 28884599]
47.
Aleppo G, Laffel LM, Ahmann AJ, et al. A practical approach to using trend arrows on the Dexcom G5 CGM system for the management of adults with diabetes. J Endocr Soc 2017;1:1445–1460 [PMC free article: PMC5760210] [PubMed: 29344577]
48.
NovoNordisk. Novo Nordisk receives FDA approval for Fiasp, a new fast-acting mealtime insulin. Available from press.novonordisk-us. com/2017-09-29-Novo-Nordisk-Receives-FDA-Approval-for-Fiasp-R-aNew-Fast-Acting-Mealtime-Insulin. Accessed 8 April 2018
49.
Dexcom. AppendixF: technical information. In Dexcom G6 Continuous Glucose Monitoring System User Guide. San Diego, CA, Dexcom, 2018, p. 289–333
50.
Garg SK, Weinzimer SA, Tamborlane WV, et al. Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery system in adolescents and adults with type 1 diabetes. Diabetes Technol Ther 2017; 19:155–163 [PMC free article: PMC5359676] [PubMed: 28134564]
51.
Weinstock RS, DuBose SN, Bergenstal RM, et al.; T1D Exchange Severe Hypoglycemia in Older Adults With Type 1 Diabetes Study Group. Risk factors associated with severe hypoglycemia in older adults with type 1 siabetes. Diabetes Care 2016;39:603–610 [PubMed: 26681721]
52.
Alitta Q, Grino M, Adjemout L, Langar A, Retornaz F, Oliver C. Overestimation of hypoglycemia diagnosis by FreeStyle Libre continuous glucose monitoring in long-term care home residents with diabetes. J Diabetes Sci Technol 2018;12:727–728 [PMC free article: PMC6154245] [PubMed: 29251062]
53.
Garber AJ, Abrahamson MJ, Barzilay JI, et al. Consensus statement by the American Association of Clinical Endocrinologists and American College of Endocrinology on the comprehensive type 2 diabetes management algorithm—2018 executive summary. Endocr Pract 2018;24:91–120 [PubMed: 29368965]
54.
Colberg SR, Sigal RJ, Yardley JE, et al. Physical activity/exercise and diabetes: a position statement of the American Diabetes Association. Diabetes Care 2016;39:2065–2079 [PMC free article: PMC6908414] [PubMed: 27926890]
55.
Riddell MC, Gallen IW, Smart CE, et al. Exercise management in type 1 diabetes: a consensus statement. Lancet Diabetes Endocrinol 2017;5:377–390 [PubMed: 28126459]
56.
Bally L, Zueger T, Pasi N, Carlos C, Paganini D, Stettler C. Accuracy of continuous glucose monitoring during differing exercise conditions. Diabetes Res Clin Pract 2016;112:1–5 [PubMed: 26739116]
57.
Moser O, Mader JK, Tschakert G, et al. Accuracy of continuous glucose monitoring (CGM) during continuous and high-intensity interval exercise in patients with type 1 diabetes mellitus. Nutrients 2016;8:E489 [PMC free article: PMC4997402] [PubMed: 27517956]
58.
Pinsker JE, Kraus A, Gianferante D, et al. Techniques for exercise preparation and management in adults with type 1 diabetes. Can J Diabetes 2016;40:503–508 [PubMed: 27212045]

Suggested citation: Hirsch IB, Battelino T, Peters AL, Chamberlain JJ, Aleppo G, Bergenstal RM. Role of Continuous Glucose Monitoring in Diabetes Treatment. Arlington, Va., American Diabetes Association, 2018

This publication has been supported by an educational grant to the American Diabetes Association from Abbott Diabetes Care.

The opinions expressed are those of the authors and do not necessarily reflect those of Abbott Diabetes Care or the American Diabetes Association. The content was developed by the authors and does not represent the policy or position of the American Diabetes Association, any of its boards or committees, or any of its journals or their editors or editorial boards.

© 2018 by American Diabetes Association. All rights reserved. None of the contents may be reproduced without the written permission of the American Diabetes Association.

Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. See http://creativecommons.org/licenses/by-nc-nd/3.0 for details.

Bookshelf ID: NBK538974PMID: 34251775DOI: 10.2337/db20181-13

Views

  • PubReader
  • Print View
  • Cite this Page
  • PDF version of this title (3.0M)

Related information

  • PMC
    PubMed Central citations
  • PubMed
    Links to PubMed

Similar articles in PubMed

See reviews...See all...

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...