Initial results from the off-label use of the SAPIEN S3 valve for percutaneous transcatheter pulmonary valve replacement: A multi-institutional experience

Catheter Cardiovasc Interv. 2019 Feb 15;93(3):455-463. doi: 10.1002/ccd.27973. Epub 2018 Dec 7.

Abstract

Objectives: To describe a multi-center experience of percutaneous transcatheter pulmonary valve replacement (TPVR) using the Edwards Sapien S3 Valve.

Background: Off-label use of the Sapien S3 valve can allow for TPVR in patients with congenital heart disease (CHD) and large diameter dysfunctional right ventricular outflow tract (RVOT). The initial experience at five centers with the SAPIEN S3 valve for TPVR is presented with a focus on procedural techniques, success, complications, and efficacy.

Methods: A retrospective review was performed of all patients with CHD and dysfunctional RVOT who underwent TPVR using Sapien S3 valve. Imaging data, procedural elements, and clinical follow-up data were collected to evaluate acute and short-term results.

Results: A total of 50 patients underwent percutaneous placement of the Sapien S3 in the pulmonary position. Of these, 38 were placed into "native RVOTs", measuring 24-32 mm in diameter, as assessed by compliant balloon sizing. In all cases, the valve was implanted after introduction and there were no cases of valve embolization. On follow up (range 1-9 months, median 3 months), no patients had significant obstruction or regurgitation through or around the valve requiring intervention. There were no procedural deaths. Major complications included severe aortic compression (n = 1) and tricuspid valve (TV) injury related to valve placement (n = 2) and prestenting (n = 1).

Conclusions: TPVR in patients with large diameter dysfunctional RVOTs can be effectively performed with the Sapien S3. All procedures were technically successful with no embolizations, no perivalvular leaks, and excellent short-term valve function. Tricuspid valve injury from implantation of an uncovered valve was a serious procedural complication.

Keywords: Sapien valve; congenital heart disease; pulmonary regurgitation; trans-catheter pulmonary valve replacement.

Publication types

  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Balloon Valvuloplasty
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Surgical Procedures / adverse effects*
  • Child
  • Female
  • Heart Defects, Congenital / physiopathology
  • Heart Defects, Congenital / surgery*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Postoperative Complications / etiology
  • Postoperative Complications / physiopathology
  • Prosthesis Design
  • Pulmonary Valve / diagnostic imaging
  • Pulmonary Valve / physiopathology
  • Pulmonary Valve / surgery*
  • Pulmonary Valve Insufficiency / diagnostic imaging
  • Pulmonary Valve Insufficiency / etiology
  • Pulmonary Valve Insufficiency / physiopathology
  • Pulmonary Valve Insufficiency / surgery*
  • Recovery of Function
  • Retrospective Studies
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • United States
  • Young Adult