Intentional Fracture of Bioprosthetic Valve Frames in Patients Undergoing Valve-in-Valve Transcatheter Pulmonary Valve Replacement

Circ Cardiovasc Interv. 2018 Aug;11(8):e006453. doi: 10.1161/CIRCINTERVENTIONS.118.006453.

Abstract

Background: Percutaneous transcatheter pulmonary valve replacement (TPVR) has good clinical and hemodynamic outcomes in treating dysfunctional bioprosthetic valves (BPV) in the pulmonary position. Valve-in-valve therapy can further decrease the inner diameter (ID), potentially resulting in patient-prosthesis mismatch in patients with smaller BPVs.

Methods and results: To evaluate feasibility and outcomes of intentional BPV fracture to enlarge the pulmonary valve orifice with TPVR, 37 patients from 13 centers who underwent TPVR with intended BPV fracture were evaluated. A control cohort (n=70) who underwent valve-in-valve TPVR without attempted fracture was evaluated. BPV was successfully fractured in 28 patients and stretched in 5 while fracture was unsuccessful in 4. A Melody valve was implanted in 25 patients with fractured/stretched frame and a Sapien (XT 3) valve in 8. Among patients whose BPV was fractured/stretched, the final ID was a median of 2 mm larger (0-6.5 mm) than the valve's true ID. The narrowest diameter after TPVR in controls was a median of 2 mm smaller ( P<0.001) than true ID. Right ventricular outflow tract gradient decreased from median 40 to 8 mm Hg in the fracture group. Cases with fracture/stretching were matched 1:1 (weight, true ID) to controls. Post-TPVR peak gradient was lower but not significant (8.3±5.2 versus 11.8±9.2 mm Hg; P=0.070). There were no fracture-related adverse events.

Conclusions: Preliminary experience shows intentional fracture of BPV frame can be useful for achieving larger ID and better hemodynamics after valve-in-valve TPVR.

Keywords: hemodynamics; prostheses and implants; pulmonary valve; stents; tetralogy of Fallot.

Publication types

  • Multicenter Study
  • Video-Audio Media

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Bioprosthesis*
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / methods
  • Child
  • Feasibility Studies
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Middle Aged
  • Preliminary Data
  • Prosthesis Design
  • Prosthesis Failure*
  • Pulmonary Valve / diagnostic imaging
  • Pulmonary Valve / physiopathology
  • Pulmonary Valve / surgery*
  • Recovery of Function
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome
  • Young Adult