3-Year Outcomes of the Edwards SAPIEN Transcatheter Heart Valve for Conduit Failure in the Pulmonary Position From the COMPASSION Multicenter Clinical Trial

JACC Cardiovasc Interv. 2018 Oct 8;11(19):1920-1929. doi: 10.1016/j.jcin.2018.06.001.

Abstract

Objectives: This study provides the 3-year follow-up results of the COMPASSION (Congenital Multicenter Trial of Pulmonic Valve Regurgitation Studying the SAPIEN Transcatheter Heart Valve) trial. Patients with moderate to severe pulmonary regurgitation and/or right ventricular outflow tract conduit obstruction were implanted with the SAPIEN transcatheter heart valve (THV).

Background: Early safety and efficacy of the Edwards SAPIEN THV in the pulmonary position have been established through a multicenter clinical trial.

Methods: Eligible patients were included if body weight was >35 kg and in situ conduit diameter was ≥16 and ≤24 mm. Adverse events were adjudicated by an independent clinical events committee. Three-year clinical and echocardiographic outcomes were evaluated in these patients.

Results: Fifty-seven of the 63 eligible patients were accounted for at the 3-year follow-up visit from a total of 69 implantations in 81 enrolled patients. THV implantation was indicated for pulmonary stenosis (7.6%), regurgitation (12.7%), or both (79.7%). Twenty-two patients (27.8%) underwent implantation of 26-mm valves, and 47 patients received 23-mm valves. Functional improvement in New York Heart Association functional class was observed in 93.5% of patients. Mean peak conduit gradient decreased from 37.5 ± 25.4 to 17.8 ± 12.4 mm Hg (p < 0.001), and mean right ventricular systolic pressure decreased from 59.6 ± 17.7 to 42.9 ± 13.4 mm Hg (p < 0.001). Pulmonary regurgitation was mild or less in 91.1% of patients. Freedom from all-cause mortality at 3 years was 98.4%. Freedom from reintervention was 93.7% and from endocarditis was 97.1% at 3 years. There were no observed stent fractures.

Conclusions: Transcatheter pulmonary valve replacement using the Edwards SAPIEN THV demonstrates excellent valve function and clinical outcomes at 3-year follow-up.

Keywords: SAPIEN THV; TPVR; conduit; pulmonary valve replacement; stent; transcatheter.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / mortality
  • Child
  • Compassionate Use Trials
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Postoperative Complications / etiology
  • Prospective Studies
  • Prosthesis Design
  • Pulmonary Valve / diagnostic imaging
  • Pulmonary Valve / physiopathology
  • Pulmonary Valve / surgery*
  • Pulmonary Valve Insufficiency / diagnostic imaging
  • Pulmonary Valve Insufficiency / mortality
  • Pulmonary Valve Insufficiency / physiopathology
  • Pulmonary Valve Insufficiency / surgery*
  • Recovery of Function
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • United States
  • Ventricular Outflow Obstruction / diagnostic imaging
  • Ventricular Outflow Obstruction / mortality
  • Ventricular Outflow Obstruction / physiopathology
  • Ventricular Outflow Obstruction / surgery*
  • Young Adult