Transcatheter Pulmonary Valve Replacement With the Melody Valve in Small Diameter Expandable Right Ventricular Outflow Tract Conduits

JACC Cardiovasc Interv. 2018 Mar 26;11(6):554-564. doi: 10.1016/j.jcin.2018.01.239.

Abstract

Objectives: This study sought to evaluate the safety, feasibility, and outcomes of transcatheter pulmonary valve replacement (TPVR) in conduits ≤16 mm in diameter.

Background: The Melody valve (Medtronic, Minneapolis, Minnesota) is approved for the treatment of dysfunctional right ventricular outflow tract (RVOT) conduits ≥16 mm in diameter at the time of implant. Limited data are available regarding the use of this device in smaller conduits.

Methods: The study retrospectively evaluated patients from 9 centers who underwent percutaneous TPVR into a conduit that was ≤16 mm in diameter at the time of implant, and reported procedural characteristics and outcomes.

Results: A total of 140 patients were included and 117 patients (78%; median age and weight 11 years of age and 35 kg, respectively) underwent successful TPVR. The median original conduit diameter was 15 (range: 9 to 16) mm, and the median narrowest conduit diameter was 11 (range: 4 to 23) mm. Conduits were enlarged to a median diameter of 19 mm (29% larger than the implanted diameter), with no difference between conduits. There was significant hemodynamic improvement post-implant, with a residual peak RVOT pressure gradient of 7 mm Hg (p < 0.001) and no significant pulmonary regurgitation. During a median follow-up of 2.0 years, freedom from RVOT reintervention was 97% and 89% at 2 and 4 years, respectively, and there were no deaths and 5 cases of endocarditis (incidence rate 2.0% per patient-year).

Conclusions: In this preliminary experience, TPVR with the Melody valve into expandable small diameter conduits was feasible and safe, with favorable early and long-term procedural and hemodynamic outcomes.

Keywords: allograft; bovine jugular vein conduit; percutaneous valve; pulmonary artery; stent; tetralogy of Fallot.

Publication types

  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Angiography
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / methods
  • Child
  • Child, Preschool
  • Feasibility Studies
  • Female
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / etiology
  • Graft Occlusion, Vascular / physiopathology
  • Graft Occlusion, Vascular / surgery*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Preliminary Data
  • Prosthesis Design
  • Pulmonary Valve / diagnostic imaging
  • Pulmonary Valve / physiopathology
  • Pulmonary Valve / surgery*
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome
  • United States
  • Ventricular Outflow Obstruction / diagnostic imaging
  • Ventricular Outflow Obstruction / etiology
  • Ventricular Outflow Obstruction / physiopathology
  • Ventricular Outflow Obstruction / surgery*
  • Young Adult