Feasibility of a Modified Cone-Beam CT Rotation Trajectory to Improve Liver Periphery Visualization during Transarterial Chemoembolization

Radiology. 2015 Dec;277(3):833-41. doi: 10.1148/radiol.2015142821. Epub 2015 May 20.

Abstract

Purpose: To compare liver coverage and tumor detectability by using preprocedural magnetic resonance (MR) images as a reference, as well as radiation exposure of cone-beam computed tomography (CT) with different rotational trajectories.

Materials and methods: Fifteen patients (nine men and six women; mean age ± standard deviation, 65 years ± 5) with primary or secondary liver cancer were retrospectively included in this institutional review board-approved study. A modified cone-beam CT protocol was used in which the C-arm rotates from +55° to -185° (open arc cone-beam CT) instead of -120° to +120° (closed arc cone-beam CT). Each patient underwent two sessions of transarterial chemoembolization between February 2013 and March 2014 with closed arc and open arc cone-beam CT (during the first and second transarterial chemoembolization sessions, respectively, as part of the institutional transarterial chemoembolization protocol). For each cone-beam CT examination, liver volume and tumor detectability were assessed by using MR images as the reference. Radiation exposure was compared by means of a phantom study. For statistical analysis, paired t tests and a Wilcoxon signed rank test were performed.

Results: Mean liver volume imaged was 1695 cm(3) ± 542 and 1857 cm(3) ± 571 at closed arc and open arc cone-beam CT, respectively. The coverage of open arc cone-beam CT was significantly higher compared with closed arc cone-beam CT (97% vs 86% of the MR imaging liver volume, P = .002). In eight patients (53%), tumors were partially or completely outside the closed arc cone-beam CT field of view. All tumors were within the open arc cone-beam CT field of view. The open arc cone-beam CT radiation exposure by means of weighted CT index was slightly lower compared with that of closed arc cone-beam CT (-5.1%).

Conclusion: Open arc cone-beam CT allowed for a significantly improved intraprocedural depiction of peripheral hepatic tumors while achieving a slight radiation exposure reduction.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Aged
  • Chemoembolization, Therapeutic / methods*
  • Cone-Beam Computed Tomography / methods*
  • Feasibility Studies
  • Female
  • Humans
  • Liver / diagnostic imaging*
  • Liver Neoplasms / therapy
  • Magnetic Resonance Imaging
  • Male
  • Radiation Exposure
  • Retrospective Studies
  • Rotation