Performance of a new blood glucose monitoring system in the hands of intended users

Diabetes Technol Ther. 2012 Sep;14(9):783-9. doi: 10.1089/dia.2012.0046. Epub 2012 Jun 29.

Abstract

Background: This study assessed the performance of a blood glucose monitoring system (BGMS) in development that uses a new generation of blood glucose test strips with capillary and venous blood in the hands of its intended users, people with diabetes and healthcare professionals (HCPs).

Subjects and methods: In total, 93 subjects ≥ 18 years old (median age, 33 years) with type 1 (78%) or type 2 (22%) diabetes participated. Untrained subjects performed self-test fingersticks using a Microlet(®)2 lancing device (Bayer HealthCare LLC, Diabetes Care, Tarrytown, NY) followed by testing of their own capillary blood on the BGMS. HCPs performed fingersticks (using a Tenderlett(®) lancing device [International Technidyne Corp., Edison, NJ]) and venipunctures on subjects and tested both capillary and venous samples from subjects on the BGMS. All BGMS results were compared with Yellow Springs Instruments (YSI) (YSI Life Sciences, Inc., Yellow Springs, OH) laboratory results. Analytical accuracy was assessed according to International Organization for Standardization (ISO) 15197:2003 guidelines (i.e., within ± 15 mg/dL or ± 20% of the YSI results for samples with glucose concentrations < 75 mg/dL and ≥ 75 mg/dL, respectively) and more stringent criteria (i.e., within ± 15 mg/dL or ± 15% of the YSI results for samples with glucose concentrations < 100 mg/dL and ≥ 100 mg/dL, respectively).

Results: Overall, 98.9% (180/182) of subject Microlet2 capillary fingerstick results, 99.5% (182/183) of HCP Tenderlett capillary fingerstick results, and 100% (186/186) of venous results met current ISO criteria and more stringent criteria. The average hematocrit was 44%, with values ranging from 32% to 52%.

Conclusions: Test results from both capillary fingerstick and venous samples with a new BGMS in development met current accuracy guidelines as well as proposed tighter criteria.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Blood Glucose / metabolism*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / standards
  • Diabetes Mellitus, Type 1 / blood*
  • Diabetes Mellitus, Type 1 / epidemiology
  • Diabetes Mellitus, Type 2 / blood*
  • Diabetes Mellitus, Type 2 / epidemiology
  • Female
  • Hematocrit / instrumentation*
  • Humans
  • Male
  • Middle Aged
  • Patient Satisfaction / statistics & numerical data
  • Reproducibility of Results
  • Surveys and Questionnaires
  • United States / epidemiology
  • Young Adult

Substances

  • Blood Glucose