Performance of the CONTOUR® TS Blood Glucose Monitoring System

J Diabetes Sci Technol. 2011 Jan 1;5(1):198-205. doi: 10.1177/193229681100500128.

Abstract

Background: Self-monitoring of blood glucose (SMBG) remains an important component of diabetes management, engendering a need for affordable blood glucose (BG) meters that are accurate, precise, and convenient. The CONTOUR® TS is a BG meter that endeavors to meet this need. It uses glucose dehydrogenase/flavin dinucleotide chemistry, automatic test strip calibration, and autocompensation for hematocrit along with the ease of use that has come to be expected of a modern meter. The objective of this clinical trial was to determine whether the CONTOUR TS system met these criteria.

Methods: The system was evaluated at a single clinical site with 106 subjects with type 1 or type 2 diabetes. Blood glucose values ranged from 60 to 333 mg/dl over all subjects. Both lay users and health care professionals (HCPs) tested the meters, with test strips from three different lots. Results were compared to a reference analyzer of verified precision and accuracy. Forty-nine of the subjects also participated in a home study of the meter. Lay users learned to use the system without assistance and were surveyed on its use at the end of the study.

Results: When used with capillary blood, both subjects and HCPs obtained results that exceeded the International Organization for Standardization 15197:2003 criteria, (i.e., ≥95% of values fell within 20% or 15 mg/dl of the laboratory value for BG levels greater than or less than 75 mg/dl, respectively). Specifically, lay users achieved 97.9% and HCPs 98.6%. When used with venous blood, 99.8% of measurements were within the criteria. All measurements for both capillary and venous blood fell into zones A or B of the Parkes error grid, deemed clinically accurate. Hematocrit was found to have no influence on BG measurements. A large majority of the subjects found the system easy to learn and to use.

Conclusions: The CONTOUR TS BG meter system gave accurate and reproducible results with both capillary and venous blood; subjects learned to use the meter system by following the user guide and quick reference guide.

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation
  • Blood Glucose Self-Monitoring / methods
  • Blood Glucose Self-Monitoring / standards
  • Calibration
  • Clinical Laboratory Techniques
  • Diabetes Mellitus, Type 1 / blood*
  • Diabetes Mellitus, Type 1 / diagnosis
  • Diabetes Mellitus, Type 1 / therapy
  • Diabetes Mellitus, Type 2 / blood*
  • Diabetes Mellitus, Type 2 / diagnosis
  • Diabetes Mellitus, Type 2 / therapy
  • Efficiency
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Physiologic / instrumentation*
  • Monitoring, Physiologic / methods*
  • Monitoring, Physiologic / standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Young Adult

Substances

  • Blood Glucose