Enhancing collection efficiency in large field of view multiphoton microscopy

J Microsc. 2011 Feb;241(2):119-24. doi: 10.1111/j.1365-2818.2010.03419.x.

Abstract

Many multiphoton imaging applications would benefit from a larger field of view; however, large field of views (>mm) require low magnification objectives which have low light collection efficiencies. We demonstrate a light collection system mounted on a low magnification objective that increases fluorescence collection by as much as 20-fold in scattering tissues. This peripheral detector results in an effective numerical aperture of collection >0.8 with a 3-4 mm field of view.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Abdominal Neoplasms / pathology
  • Animals
  • Brain / cytology
  • Carcinoma / pathology
  • Mice
  • Microscopy, Fluorescence / methods*
  • Pathology / methods*
  • Spinal Cord / cytology