Independent monitor unit calculation for intensity modulated radiotherapy using the MIMiC multileaf collimator

Med Phys. 2002 Sep;29(9):2041-51. doi: 10.1118/1.1500397.

Abstract

A self-consistent monitor unit (MU) and isocenter point-dose calculation method has been developed that provides an independent verification of the MU for intensity modulated radiotherapy (IMRT) using the MIMiC (Nomos Corporation) multileaf collimator. The method takes into account two unique features of IMRT using the MIMiC: namely the gantry-dynamic arc delivery of intensity modulated photon beams and the slice-by-slice dose delivery for large tumor volumes. The method converts the nonuniform beam intensity planned at discrete gantry angles of 5 degrees or 10 degrees into conventional nonmodulated beam intensity apertures of elemental arc segments of 1 degree. This approach more closely simulates the actual gantry-dynamic arc delivery by MIMiC. Because each elemental arc segment is of uniform intensity, the MU calculation for an IMRT arc is made equivalent to a conventional arc with gantry-angle dependent beam apertures. The dose to the isocenter from each 1 degree elemental arc segment is calculated by using the Clarkson scatter summation technique based on measured tissue-maximum-ratio and output factors, independent of the dose calculation model used in the IMRT planning system. For treatments requiring multiple treatment slices, the MU for the arc at each treatment slice takes into account the MU, leakage and scatter doses from other slices. This is achieved by solving a set of coupled linear equations for the MUs of all involved treatment slices. All input dosimetry data for the independent MU/isocenter point-dose calculation are measured directly. Comparison of the MU and isocenter point dose calculated by the independent program to those calculated by the Corvus planning system and to direct measurements has shown good agreement with relative difference less than +/-3%. The program can be used as an independent initial MU verification for IMRT plans using the MIMiC multileaf collimators.

Publication types

  • Comparative Study
  • Evaluation Study
  • Validation Study

MeSH terms

  • Algorithms*
  • Computer Simulation
  • Equipment Design
  • Models, Biological
  • Photons
  • Quality Control
  • Radiometry / instrumentation
  • Radiometry / methods*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / instrumentation
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Conformal / instrumentation
  • Radiotherapy, Conformal / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity