A comparative evaluation of the diagnostic efficacy of film and digital sensors for detection of simulated periapical lesions

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 Jul;92(1):93-7. doi: 10.1067/moe.2001.115974.

Abstract

Objective: The purpose of this study was to evaluate the diagnostic efficacy of Ektaspeed Plus film (EPF), a charge-coupled device (CCD), and photostimulable phosphor (PSP)-based digital images for detection of simulated periapical lesions.

Study design: Lesions were simulated in the periapical areas of 24 human mandibular sections invested in acrylic using burs of sizes #1, 2, 4, and 6 and imaged using EPF, CCD, and PSP sensors. Percent correct response scores, sensitivity, and specificity values were computed for all variables. Repeated-measures analysis of variance and post hoc testing were performed to determine the effects of imaging modality, observer, and lesion sizes with respect to lesion detection.

Results: EPF displayed the highest sensitivity and specificity, followed by PSP and CCD images (P <.001). Percent correct score was the highest for 3 of 4 observers when EPF was used. Analysis of variance revealed significance (P <.001) with respect to all variables. Observers with experience in digital image-viewing performed better than those without such experience (P <.001). Intraobserver agreement was fair (kappa = 0.5).

Conclusions: EPF outperformed CCD and PSP images when observers could manipulate image characteristics.

Publication types

  • Comparative Study

MeSH terms

  • Analysis of Variance
  • Humans
  • Mandible
  • Observer Variation
  • Periapical Diseases / diagnostic imaging*
  • Radiographic Image Enhancement / instrumentation
  • Radiographic Image Enhancement / methods
  • Radiography, Dental, Digital / instrumentation*
  • Radiography, Dental, Digital / methods
  • Sensitivity and Specificity
  • Statistics as Topic
  • X-Ray Film*