One-step immunoassay of C-reactive protein using droplet microfluidics

Lab Chip. 2016 Nov 1;16(22):4359-4365. doi: 10.1039/c6lc01121g.

Abstract

We present a wash-free high-sensitivity immunoassay of C-reactive proteins with droplet microfluidics. Microbeads are encapsulated within droplets for the immunoassay, and the droplets are scanned by a fluorescence detection platform to quantify the amount of proteins captured on the microbeads. The limit of detection determined by our platform is 0.01 μg mL-1, which is ten times more sensitive than conventional high-sensitivity C-reactive protein assays. With the decrease in diffusion distance within droplets, the immunoassay requires only half of the time required for similar conventional approaches. This approach for carrying out immunoassays can potentially be applied to other biomarkers beyond C-reactive proteins.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • C-Reactive Protein / analysis*
  • Immunoassay / instrumentation*
  • Lab-On-A-Chip Devices*
  • Limit of Detection
  • Spectrometry, Fluorescence

Substances

  • C-Reactive Protein