Image comparative assessment using iterative reconstructions: clinical comparison of low-dose abdominal/pelvic computed tomography between adaptive statistical, model-based iterative reconstructions and traditional filtered back projection in 65 patients

Invest Radiol. 2014 Apr;49(4):209-16. doi: 10.1097/RLI.0000000000000017.

Abstract

Objectives: The objective of this study was to compare image quality (objective and subjective parameters) and confidence in lesion detection between 3 image reconstruction algorithms in computed tomographic (CT) examinations of the abdomen/pelvis.

Materials and methods: This prospective institutional review board-approved study included 65 patients (mean [SD] age, 71.3 ± 9 years; mean [SD] body mass index, 24.4 [4.8] kg) who underwent routine CT examinations of the abdomen/pelvis followed immediately by 2 low-dose scans. Raw data sets were reconstructed by using filtered back projection (FBP), adaptive statistical iterative reconstruction (ASIR), and a model-based iterative reconstruction (MBIR). Measurements of objective noise and CT numbers were compared using repeated-measures analysis of variance. Six subjective image quality parameters were scored. Diagnostic confidence and accuracy in detection of various elementary lesions were performed.

Results: Objectively, mean image noise for MBIR was significantly superior at all dose levels (P < 0.001). Subjectively, standard-dose ASIR and low-dose MBIR scans were better than standard-dose FBP scan in all parameters assessed (P < 0.05). Low-dose MBIR scans were comparable with standard-dose ASIR scans in all parameters except at noise index of 70 (approximately 85% dose reduction), where, in this case, the detection of liver lesions less than 5 mm were rated inferior (P < 0.05) with diagnostic accuracy reducing to 77.4%.

Conclusions: Low-dose MBIR scan shows superior objective noise reduction compared with standard-dose FBP and ASIR. Subjectively, low-dose MBIR scans at 76% dose reduction were also superior compared with standard-dose FBP and ASIR. However, at dose reductions of 85%, small liver lesions may be missed.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Abdominal Neoplasms / diagnostic imaging*
  • Aged
  • Aged, 80 and over
  • Algorithms*
  • Computer Simulation
  • Data Interpretation, Statistical
  • Female
  • Humans
  • Male
  • Middle Aged
  • Models, Biological
  • Models, Statistical
  • Pelvic Neoplasms / diagnostic imaging*
  • Pelvis / diagnostic imaging*
  • Radiographic Image Enhancement / methods
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Radiography, Abdominal / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*