Evaluation of an improved orthognathic articulator system: 1. Accuracy of cast orientation

Int J Oral Maxillofac Surg. 2012 Feb;41(2):150-4. doi: 10.1016/j.ijom.2011.09.019. Epub 2011 Nov 1.

Abstract

A systematic study was carried out using plastic model skulls to quantify the accuracy of the transfer of face bow registration to the articulator. A standard Dentatus semi-adjustable articulator system was compared to a purpose built orthognathic articulator system by measuring the maxillary occlusal plane angles of plastic model skulls and of dental casts mounted on the two different types of articulators. There was a statistically significant difference between the two systems; the orthognathic system showed small random errors, but the standard system showed systematic errors of up to 28°.