Intraoperative fracture of phacoemulsification sleeve

BMC Ophthalmol. 2010 Nov 30:10:29. doi: 10.1186/1471-2415-10-29.

Abstract

Background: We describe a case of intraoperative fracture of phacoemulsification sleeve during phacoemulsification surgery.

Case presentation: Phacoemulsification surgery was performed in the left eye of a 58-year-old lady with grade II nuclear sclerosis & grade I cortical cataract. Towards the end of quadrant removal, there was anterior chamber instability with impaired followability of nuclear fragments. The distal part of the fractured sleeve remained inside the anterior chamber upon removal of the phacoemulsification probe. The retained sleeve was retrieved with a pair of forceps through the corneal incision site, which did not require widening. There was no missing fragments retained intraocularly and the patient had an uneventful recovery with vision of 20/25 at three months post-operatively.

Conclusion: Phacoemulsification sleeve fracture is an uncommon complication. With early identification of this condition and proper management, major complications can be avoided.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Anterior Chamber / physiopathology
  • Cataract Extraction / methods*
  • Cornea / metabolism
  • Cornea / pathology
  • Device Removal / instrumentation
  • Device Removal / methods
  • Equipment Failure
  • Eyeglasses
  • Female
  • Humans
  • Intraoperative Complications* / physiopathology
  • Lens Nucleus, Crystalline / pathology*
  • Phacoemulsification / adverse effects*
  • Phacoemulsification / instrumentation*
  • Sclerosis
  • Sodium Chloride / pharmacokinetics
  • Surgical Instruments
  • Treatment Outcome
  • Visual Acuity

Substances

  • Sodium Chloride