Effects of scan circle displacement in optical coherence tomography retinal nerve fibre layer thickness measurement: a RNFL modelling study

Eye (Lond). 2009 Jun;23(6):1436-41. doi: 10.1038/eye.2008.258. Epub 2008 Sep 19.

Abstract

Objective: To study the effect of optical coherence tomography (OCT) scan circle displacement on retinal nerve fibre layer (RNFL) measurement errors using cubic spline models.

Methods: Forty-nine normal subjects were included in the analysis. In one randomly selected eye in each subject, RNFL thickness around the optic disc was measured by taking 16 circular scans of different sizes (scan radius ranged from 1 to 2.5 mm). The RNFL profile in each eye was constructed with a mathematical model using a smoothing spline approximation. Scan circle (diameter 3.4 mm) RNFL measurements (total average, superior, nasal, inferior, and temporal RNFL thicknesses) obtained from eight directions (superior, superonasal, nasal, inferonasal, inferior, inferotemporal, temporal, and superotemporal) displaced at different distances (0.1, 0.2, 0.3, 0.4, 0.5, 0.6, and 0.7 mm) from the disc centre were then computed by a computer program and compared to the 'reference standard' where the scan circle is centred at the optic disc. RNFL measurement error was calculated as the absolute of (RNFL thickness(displaced) - RNFL thickness(reference standard)).

Results: The respective mean average, superior, nasal, inferior, and temporal RNFL measurement errors were 2.3+/-2.0, 4.9+/-4.5, 4.1+/-3.8, 6.2+/-7.6, and 3.8+/-3.5 microm upon 0.1 mm scan circle displacement, and 12.1+/-11.4, 27.8+/-18.4, 21.7+/-18.6, 34.8+/-22.9, and 15.2+/-10.7 microm upon 0.7 mm scan circle displacement. Significant differences of average and quadrant RNFL thicknesses were evident between centred and displaced scan circle measurements (all with P<0.001). RNFL measurement error increased in a monotonic fashion with increasing distance away from the disc and the change was direction-dependent.

Conclusions: RNFL measurement error varies with the direction and distance of scan displacement. The superior and the inferior RNFL measurements are most vulnerable to scan displacement errors, whereas the average RNFL thickness is the least susceptible. Obtaining a well-centred scan is essential for reliable RNFL measurement in OCT.

MeSH terms

  • Adult
  • Female
  • Humans
  • Imaging, Three-Dimensional / methods
  • Male
  • Middle Aged
  • Models, Biological
  • Nerve Fibers*
  • Optic Disk / anatomy & histology*
  • Retinal Ganglion Cells / cytology*
  • Tomography, Optical Coherence*
  • Young Adult