Rate responsive pacing with a minute ventilation sensing pacemaker during pregnancy and delivery

Pacing Clin Electrophysiol. 1990 Feb;13(2):158-63. doi: 10.1111/j.1540-8159.1990.tb05065.x.

Abstract

A minute ventilation sensing rate responsive pacemaker was implanted in a 29-week pregnant woman with symptomatic complete atrioventricular (AV) block under echocardiographic and electrocardiographic (ECG) guidance. Satisfactory rate responses during a submaximal treadmill exercise test and daily activities were achieved. The course of pregnancy and cesarean section is discussed. During the cesarean section and after the delivery of the baby, changes in the ventilator settings confirmed that the pacing rate was closely correlated with both the tidal volume (r = 0.94, P less than 0.01) and the respiratory rate (r = 0.93, P less than 0.05).

Publication types

  • Case Reports

MeSH terms

  • Activities of Daily Living
  • Adult
  • Blood Pressure / physiology
  • Cardiac Pacing, Artificial / methods*
  • Cesarean Section*
  • Electrocardiography, Ambulatory
  • Equipment Design
  • Female
  • Heart Block / complications
  • Heart Block / physiopathology
  • Heart Block / therapy*
  • Heart Rate, Fetal / physiology
  • Humans
  • Pacemaker, Artificial*
  • Physical Exertion
  • Pregnancy
  • Pregnancy Complications* / physiopathology
  • Prenatal Diagnosis
  • Respiration / physiology*
  • Ultrasonography