Interface for capillary electrophoresis coupled with inductively coupled plasma atomic emission spectrometry

J Chromatogr A. 1999 Aug 20;853(1-2):141-9. doi: 10.1016/s0021-9673(99)00713-x.

Abstract

A modified concentric nebulizer was used as the interface to couple capillary electrophoresis (CE) to inductively coupled plasma atomic emission spectrometry (ICP-AES). The CE capillary replaces the central tube of the concentric nebulizer. The tip of the nebulizer tapers slowly to allow uncertainty in the position of the capillary. A platinum wire was inserted into the CE capillary to provide electrical connection to the CE power supply. pH changes inside the capillary due to electrolysis of the background buffer electrolyte was small and has minimal effects on the CE separation. The peak broadening effects due to the nebulizing gas flow, however, were significant. Resolution decreases quickly when the flow-rate of the carrier gas increases. Sample stacking technique was used to improve the resolution of species of opposite charge, e.g., Cr(VI) vs. Cr(III) ions. Detection limit of Cr based on peak area is approximately 10 ppb for the CE-ICP-AES system.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromium / analysis*
  • Electrodes
  • Electrolysis
  • Electrophoresis, Capillary / instrumentation
  • Electrophoresis, Capillary / methods*
  • Nebulizers and Vaporizers
  • Spectrum Analysis
  • Time Factors

Substances

  • Chromium