Atrial fibrillation detection and R-wave synchronization by Metrix implantable atrial defibrillator: implications for long-term efficacy and safety. The Metrix Investigators

Circulation. 1999 Mar 23;99(11):1446-51. doi: 10.1161/01.cir.99.11.1446.

Abstract

Background: The long-term efficacy of atrial fibrillation (AF) detection and R-wave synchronization are critical safety requirements for the development of an implantable atrial defibrillator (IAD) for treatment of AF.

Methods and results: The long-term efficacy of the Metrix IAD for AF detection and R-wave synchronization was tested in 51 patients. The mean duration of follow-up was 259+/-138 days (72 to 613 days). AF detection tests were performed 2240 times during observed operation with 100% specificity and 92.3% sensitivity for differentiation between sinus rhythm and AF; 2219 episodes and their electrograms stored in the device during AF detection were analyzed. The positive predictive value of the AF detection algorithm was 97.4% (lower 95% confidence limit [CL], 94.5%) in the out-of-hospital setting. A total of 242 435 R waves were analyzed for R-wave synchronization. Of these, 49% were marked for synchronized shock delivery, 82% of sinus rhythm and 36% of AF R waves, respectively. All shock markers were properly synchronized and within the R wave (overall synchronization accuracy, 100%; lower 95% CL, 99.999%). Overall, 3719 shocks have been delivered via the IAD with no instance of unsynchronized shock delivery or any episode of proarrhythmia. The observed proarrhythmic risk was 0%, with an estimated maximum proarrhythmic risk of 0.084% per shock (95% upper CL).

Conclusions: The Metrix IAD can appropriately detect AF with a high specificity and sensitivity and reliably synchronize within a suitable R wave for shock delivery to minimize the risk of ventricular proarrhythmia.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Aged
  • Algorithms
  • Atrial Fibrillation / diagnosis*
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / surgery*
  • Cohort Studies
  • Defibrillators, Implantable* / adverse effects
  • Equipment Design
  • Female
  • Heart Conduction System / physiopathology
  • Humans
  • Male
  • Middle Aged
  • Risk
  • Safety
  • Sensitivity and Specificity
  • Ventricular Fibrillation / etiology
  • Ventricular Fibrillation / prevention & control